5I08
Prefusion structure of a human coronavirus spike protein
ELECTRON MICROSCOPY
Sample |
---|
HKU1 spike with attached foldon domain and mutated furin-cleavage site |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | HOMEMADE PLUNGER |
Cryogen Name | ETHANE |
Sample Vitrification Details | 3 uL sample was applied to grid, blotted, and plunged into liquid ethane. |
Staining Type | NEGATIVE |
Staining Material | uranyl formate |
Staining Details | 3 uL sample was applied to grid for 30 seconds and then blotted. Grids were stained with 3 uL 1% uranyl formate for 60 seconds followed by blotting. |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 31435 |
Reported Resolution (Å) | 4.04 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C3 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | AB INITIO MODEL | ||||
Refinement Target | EMRinger | ||||
Overall B Value | 117 | ||||
Fitting Procedure | |||||
Details | Model building and refinement were conducted using a combination of software programs. |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K2 SUMMIT (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 57 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 1000 |
Maximum Defocus (nm) | 3500 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 22500 |
Calibrated Magnification | 22500 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | DoG Picker | |
PARTICLE SELECTION | FindEM | |
IMAGE ACQUISITION | Leginon | |
CTF CORRECTION | CTFFIND3 | |
MODEL FITTING | MODELLER | |
MODEL FITTING | Rosetta | 2015.19 |
MODEL FITTING | Coot | 0.8.2 |
INITIAL EULER ASSIGNMENT | RELION | 1.3 |
FINAL EULER ASSIGNMENT | RELION | 1.4b1 |
CLASSIFICATION | IMAGIC | |
RECONSTRUCTION | RELION | 1.4b1 |
MODEL REFINEMENT | Rosetta | 2015.19 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | 39164 | 2188 particles were selected from a subset of the data using DoG Picker. These particles were used to generate a 3D model from which back projections were derived. Back projection images were used as templates for picking particles from the entire dataset using FindEM. |