blotting at room temperature with sample at room temperature
3D Reconstruction
Reconstruction Method
SINGLE PARTICLE
Number of Particles
48271
Reported Resolution (Å)
6.9
Resolution Method
FSC 0.5 CUT-OFF
Other Details
Projection matching was done with Fourier ring correlation, model-based masking and SSNR weighting over an 80-6 angstrom resolution range. The final r ...
Projection matching was done with Fourier ring correlation, model-based masking and SSNR weighting over an 80-6 angstrom resolution range. The final refinement steps used an angular step of 2.5 degrees and each of the 48,000 particles was matched to the best two projection classes (1353). In total, 45,000 particles were used in the final reconstruction and the 3D map was amplitude corrected then Gaussian low-pass filtered with a Fourier half-width of 0.12.
Refinement Type
Symmetry Type
POINT
Point Symmetry
D8
Map-Model Fitting and Refinement
Id
1
Refinement Space
REAL
Refinement Protocol
FLEXIBLE FIT
Refinement Target
Cross-correlation coefficient
Overall B Value
Fitting Procedure
Details
METHOD--Flexible fitting, local refinement REFINEMENT PROTOCOL--rigid body
first, and then used individual helices DETAILS--Chimera was used to do in ...
METHOD--Flexible fitting, local refinement REFINEMENT PROTOCOL--rigid body
first, and then used individual helices DETAILS--Chimera was used to do initial
domain fitting with Apaf-1 (1Z6T) and CED-4 (2A5Y)domains. Modeller and Flex-Em
were then used to do local refinement of homology models of the various domains.
Data Acquisition
Detector Type
GENERIC TVIPS (4k x 4k)
Electron Dose (electrons/Å**2)
20
Imaging Experiment
1
Date of Experiment
2009-09-15
Temperature (Kelvin)
93
Microscope Model
FEI TECNAI F20
Minimum Defocus (nm)
1500
Maximum Defocus (nm)
3000
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS
2
Imaging Mode
BRIGHT FIELD
Specimen Holder Model
GATAN LIQUID NITROGEN
Nominal Magnification
50000
Calibrated Magnification
50000
Source
FIELD EMISSION GUN
Acceleration Voltage (kV)
160
Imaging Details
actual magnification at the ccd 87000, camera pixel size 15um, 1.72 angstrom per pixel, data collected semi-automatically with EMTools (TVIPS)