Find PDB structures and Computed Structure Models (CSM) by combining queries from tools in this suite: Attribute Search, Sequence Similarity, Sequence Motif, 3D Similarity, and 3D Motif with 'AND' logic.
3-4 ul of sample blotted for 3.5 seconds with -15 blot force on FEI Vitrobot mark IV.
3D Reconstruction
Reconstruction Method
SINGLE PARTICLE
Number of Particles
23721
Reported Resolution (Å)
2.75
Resolution Method
FSC 0.143 CUT-OFF
Other Details
Icosahedral symmetry applied to final reconstruction. Final maps were post-processed with an inverse B-factor of -146.7 Angstrom squared.
Refinement Type
Symmetry Type
POINT
Point Symmetry
I
Map-Model Fitting and Refinement
Id
1 (1HXS)
Refinement Space
REAL
Refinement Protocol
RIGID BODY FIT
Refinement Target
Cross-correlation coefficient
Overall B Value
Fitting Procedure
Details
Initial model was rigid body fitted using UCSF chimera and Coot. Global minimization and B-factor refinement
was performed in real space using pheni ...
Initial model was rigid body fitted using UCSF chimera and Coot. Global minimization and B-factor refinement
was performed in real space using phenix_real.space.refine.
Data Acquisition
Detector Type
FEI FALCON III (4k x 4k)
Electron Dose (electrons/Å**2)
35.49
Imaging Experiment
1
Date of Experiment
Temperature (Kelvin)
Microscope Model
FEI TITAN KRIOS
Minimum Defocus (nm)
800
Maximum Defocus (nm)
2900
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS
2.7
Imaging Mode
DARK FIELD
Specimen Holder Model
FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification
75000
Calibrated Magnification
129629
Source
FIELD EMISSION GUN
Acceleration Voltage (kV)
300
Imaging Details
Calibrated pixel size was 1.08 A/pix.
EM Software
Task
Software Package
Version
PARTICLE SELECTION
cryoSPARC
4.2.1
CTF CORRECTION
cryoSPARC
4.2.1
MODEL FITTING
UCSF Chimera
1.16
MODEL FITTING
Coot
0.9.6
INITIAL EULER ASSIGNMENT
cryoSPARC
4.2.1
FINAL EULER ASSIGNMENT
cryoSPARC
4.2.1
CLASSIFICATION
cryoSPARC
4.2.1
RECONSTRUCTION
cryoSPARC
4.2.1
MODEL REFINEMENT
PHENIX
1.20.1_4487:
Image Processing
CTF Correction Type
CTF Correction Details
Number of Particles Selected
Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION
159830
Particle picking performed in cryoSPARC using the blob picker.