8FF3
Amyloid-beta (1-40) fibrils derived from familial Dutch-type CAA patient (population B)
ELECTRON MICROSCOPY
Starting Model(s)
Initial Refinement Model(s) | |||
---|---|---|---|
Type | Source | Accession Code | Details |
experimental model | PDB | 6W0O |
Refinement
RMS Deviations | |
---|---|
Key | Refinement Restraint Deviation |
f_dihedral_angle_d | 8.701 |
f_angle_d | 0.417 |
f_chiral_restr | 0.056 |
f_bond_d | 0.002 |
f_plane_restr | 0.001 |
Sample |
---|
Fibrillar assembly of human Amyloid-beta 1-40 from a familial Dutch-type cerebral amyloid angiopathy patient |
Specimen Preparation | |
---|---|
Sample Aggregation State | FILAMENT |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details |
3D Reconstruction | |
---|---|
Reconstruction Method | HELICAL |
Number of Particles | 73287 |
Reported Resolution (Å) | 3.09 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | HELICAL |
Axial Symmetry | C1 |
Axial Rise | 2.44 |
Angular Rotation | 179.75 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (6W0O) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | AB INITIO MODEL | ||||
Refinement Target | |||||
Overall B Value | 71.7 | ||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | FEI FALCON III (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 56.92 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TALOS ARCTICA |
Minimum Defocus (nm) | 600 |
Maximum Defocus (nm) | 1400 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | |
Nominal Magnification | 92000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 200 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
IMAGE ACQUISITION | EPU | |
CTF CORRECTION | Gctf | 1.06 |
CTF CORRECTION | RELION | 3.1.1 |
MODEL FITTING | Coot | |
INITIAL EULER ASSIGNMENT | RELION | 3.1.1 |
FINAL EULER ASSIGNMENT | RELION | 3.1.1 |
CLASSIFICATION | RELION | 3.1.1 |
RECONSTRUCTION | RELION | 3.1.1 |
MODEL REFINEMENT | PHENIX | 1.19.1 |
MODEL REFINEMENT | ISOLDE | 1.4 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | Per-micrograph correction via Gctf. Aberration, magnification, and per-particle corrections via RELION. | 6946 | Manual helical picking |