7P1T
Cryo-EM structure of encapsulin from Mycobacterium tuberculosis
ELECTRON MICROSCOPY
Starting Model(s)
Initial Refinement Model(s) | |||
---|---|---|---|
Type | Source | Accession Code | Details |
experimental model | PDB | 619G |
Sample |
---|
encapsulin shell from Mycobacterium tuberculosis |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 57805 |
Reported Resolution (Å) | 2.29 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | I |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (619G) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | AB INITIO MODEL | ||||
Refinement Target | multiple | ||||
Overall B Value | 34.36 | ||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K3 (6k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 40 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 700 |
Maximum Defocus (nm) | 2000 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 163000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | RELION | 3.1 |
IMAGE ACQUISITION | EPU | |
CTF CORRECTION | Gctf | |
MODEL FITTING | PHENIX | 1.18.2 |
MODEL FITTING | Coot | 0.9.5 |
MODEL REFINEMENT | PHENIX | 1.18.2 |
MODEL REFINEMENT | Coot | 0.9.5 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | 89072 |