☰ Navigation Tabs
Cryo-EM structure of conjugative pili from carbapenem-resistant Klebsiella pneumoniae
Refinement RMS Deviations Key Refinement Restraint Deviation f_dihedral_angle_d 18.806 f_angle_d 0.758 f_chiral_restr 0.04 f_bond_d 0.008 f_plane_restr 0.006
Specimen Preparation Sample Aggregation State FILAMENT Vitrification Instrument Cryogen Name ETHANE Sample Vitrification Details
3D Reconstruction Reconstruction Method HELICAL Number of Particles 163128 Reported Resolution (Å) 3.9 Resolution Method FSC 0.143 CUT-OFF Other Details Refinement Type Symmetry Type HELICAL Axial Symmetry C1 Axial Rise 2.7 Angular Rotation 77.6
Map-Model Fitting and Refinement Id 1 Refinement Space Refinement Protocol Refinement Target Overall B Value Fitting Procedure Details
Data Acquisition Detector Type GATAN K3 (6k x 4k) Electron Dose (electrons/Å**2) 55
Imaging Experiment 1 Date of Experiment Temperature (Kelvin) Microscope Model FEI TITAN KRIOS Minimum Defocus (nm) Maximum Defocus (nm) Minimum Tilt Angle (degrees) Maximum Tilt Angle (degrees) Nominal CS Imaging Mode BRIGHT FIELD Specimen Holder Model Nominal Magnification Calibrated Magnification Source FIELD EMISSION GUN Acceleration Voltage (kV) 300 Imaging Details
EM Software Task Software Package Version IMAGE ACQUISITION Latitude CTF CORRECTION RELION 3.0 MODEL FITTING RosettaEM INITIAL EULER ASSIGNMENT RELION RECONSTRUCTION RELION MODEL REFINEMENT Coot MODEL REFINEMENT PHENIX