Find PDB structures and Computed Structure Models (CSM) by combining queries from tools in this suite: Attribute Search, Sequence Similarity, Sequence Motif, 3D Similarity, and 3D Motif with 'AND' logic.
Amyloid fibril of an antibody lambda 1 light chain
Specimen Preparation
Sample Aggregation State
FILAMENT
Vitrification Instrument
GATAN CRYOPLUNGE 3
Cryogen Name
ETHANE
Sample Vitrification Details
blotted from the backside for 4 s before plunging
3D Reconstruction
Reconstruction Method
HELICAL
Number of Particles
32677
Reported Resolution (Å)
3.3
Resolution Method
FSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry Type
HELICAL
Axial Symmetry
C1
Axial Rise
4.81
Angular Rotation
0.58
Map-Model Fitting and Refinement
Id
1
Refinement Space
REAL
Refinement Protocol
OTHER
Refinement Target
REAL-SPACE (WEIGHTED MAP SUM AT ATOM CENTERS)
Overall B Value
111.4
Fitting Procedure
Details
Refinement strategy included global minimization and local grid search and ADP were refined. Secondary structure restraints and NCS were applied durin ...
Refinement strategy included global minimization and local grid search and ADP were refined. Secondary structure restraints and NCS were applied during refinement.
Data Acquisition
Detector Type
GATAN K2 SUMMIT (4k x 4k)
Electron Dose (electrons/Å**2)
32
Imaging Experiment
1
Date of Experiment
Temperature (Kelvin)
Microscope Model
FEI TITAN KRIOS
Minimum Defocus (nm)
Maximum Defocus (nm)
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS
2.7
Imaging Mode
BRIGHT FIELD
Specimen Holder Model
FEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification
Calibrated Magnification
Source
FIELD EMISSION GUN
Acceleration Voltage (kV)
300
Imaging Details
EM Software
Task
Software Package
Version
IMAGE ACQUISITION
SerialEM
CTF CORRECTION
Gctf
1.06
MODEL FITTING
Warp
8.0
MODEL FITTING
Coot
0.8.9
INITIAL EULER ASSIGNMENT
RELION
2.1.0
CLASSIFICATION
RELION
2.1.0
RECONSTRUCTION
RELION
2.1.0
MODEL REFINEMENT
PHENIX
1.14.3260
Image Processing
CTF Correction Type
CTF Correction Details
Number of Particles Selected
Particle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION
CTF was estimated from the non-dose-weighted micrographs
119395
manual selection. Sigma contrast 3, lowpass filter 20 A