6EK5
Near-atomic resolution structure of a plant geminivirus determined by electron cryo-microscopy.
ELECTRON MICROSCOPY
Sample |
---|
African cassava mosaic virus - [West Kenya 844] |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK IV |
Cryogen Name | ETHANE |
Sample Vitrification Details | Samples (3 ?l) were applied to the glow discharged grids, incubated for 60 s on the grid, blotted and plunge frozen in liquid ethane using a Vitrobot ... |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 24451 |
Reported Resolution (Å) | 4.2 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | D5 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | FLEXIBLE FIT | ||||
Refinement Target | |||||
Overall B Value | 139 | ||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | FEI FALCON II (4k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 25 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | |
Maximum Defocus (nm) | |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 0.01 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 94000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | RELION | 1.4 |
CTF CORRECTION | RELION | 1.4 |
MODEL FITTING | Coot | |
MODEL REFINEMENT | PHENIX | |
INITIAL EULER ASSIGNMENT | RELION | 1.4 |
FINAL EULER ASSIGNMENT | RELION | 1.4 |
CLASSIFICATION | RELION | 1.4 |
RECONSTRUCTION | RELION | 1.4 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | 141141 | For the initial analysis, ca. 2,000 particle images were selected manually with e2boxer and extracted and normalized with relion 1.4 followed by 2D-classification and ctf-correction. Three characteristic class averages that resolved the two parts of the twin particle (side views and intermediate views) were selected as references for template-dependent automatic particle picking in Relion. Particle images were extracted at the determined coordinates with a box size of 300 x 300 Px and normalized for their grey value distribution followed by 2D-classification. Some of the 2D-classes showed disconnected twin particles or single capsids. These classes probably represented either adjacent capsids from different twin particles or incorrectly centered particles and were excluded from the subsequent analysis. 141141 particles is the number of automatically selected particles; 69685 were retained for he subsequent processing |