6X2L

hEAAT3-IFS-Na


ELECTRON MICROSCOPY

Refinement

RMS Deviations
KeyRefinement Restraint Deviation
f_dihedral_angle_d18.121
f_angle_d0.97
f_chiral_restr0.055
f_bond_d0.009
f_plane_restr0.007
Sample
inward facing hEAAT3 trimer
Sample Components
inward facing hEAAT3 trimer
Specimen Preparation
Sample Aggregation StatePARTICLE
Vitrification InstrumentFEI VITROBOT MARK IV
Cryogen NameETHANE
Sample Vitrification Detailsblot 3s
3D Reconstruction
Reconstruction MethodSINGLE PARTICLE
Number of Particles554920
Reported Resolution (Å)2.85
Resolution MethodFSC 0.143 CUT-OFF
Other Details
Refinement Type
Symmetry TypePOINT
Point SymmetryC3
Map-Model Fitting and Refinement
Id1
Refinement Space
Refinement Protocol
Refinement Target
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector TypeGATAN K3 BIOQUANTUM (6k x 4k)
Electron Dose (electrons/Å**2)60
Imaging Experiment1
Date of Experiment
Temperature (Kelvin)
Microscope ModelFEI TITAN KRIOS
Minimum Defocus (nm)500
Maximum Defocus (nm)1500
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS2.7
Imaging ModeBRIGHT FIELD
Specimen Holder ModelFEI TITAN KRIOS AUTOGRID HOLDER
Nominal Magnification105000
Calibrated Magnification
SourceFIELD EMISSION GUN
Acceleration Voltage (kV)300
Imaging Details
EM Software
TaskSoftware PackageVersion
PARTICLE SELECTIONRELION3.0
IMAGE ACQUISITIONSerialEM
CTF CORRECTIONRELION3.0
INITIAL EULER ASSIGNMENTRELION3.0
FINAL EULER ASSIGNMENTRELION3.0
RECONSTRUCTIONRELION3.0
Image Processing
CTF Correction TypeCTF Correction DetailsNumber of Particles SelectedParticle Selection Details
PHASE FLIPPING AND AMPLITUDE CORRECTION1717667