6W1J
Cryo-EM structure of 5HT3A receptor in presence of Alosetron
ELECTRON MICROSCOPY
Sample |
---|
Serotonin receptor |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | FEI VITROBOT MARK I |
Cryogen Name | ETHANE |
Sample Vitrification Details | blot for 2.5 seconds |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 42065 |
Reported Resolution (Å) | 2.92 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C5 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 | ||||
Refinement Space | REAL | ||||
Refinement Protocol | OTHER | ||||
Refinement Target | |||||
Overall B Value | 30 | ||||
Fitting Procedure | |||||
Details |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K3 (6k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 50 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | |
Maximum Defocus (nm) | |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | |
Nominal Magnification | 105000 |
Calibrated Magnification | |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | RELION | 3.1 |
CTF CORRECTION | Gctf | 1.06 |
MODEL FITTING | Coot | 0.8.9 |
INITIAL EULER ASSIGNMENT | RELION | 3.1 |
FINAL EULER ASSIGNMENT | RELION | 3.1 |
CLASSIFICATION | RELION | 3.1 |
RECONSTRUCTION | RELION | 3.1 |
MODEL REFINEMENT | PHENIX | 1.13-2998 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | 568452 |