☰ Navigation Tabs
Structure of NLRC4 CARD filament
Refinement RMS Deviations Key Refinement Restraint Deviation f_dihedral_angle_d 11.782 f_angle_d 0.364 f_chiral_restr 0.037 f_plane_restr 0.003 f_bond_d 0.002
Sample NLRC4 CARD filament
Specimen Preparation Sample Aggregation State FILAMENT Vitrification Instrument FEI VITROBOT MARK III Cryogen Name ETHANE Sample Vitrification Details
3D Reconstruction Reconstruction Method HELICAL Number of Particles 120602 Reported Resolution (Å) 3.3 Resolution Method FSC 0.143 CUT-OFF Other Details Refinement Type Symmetry Type HELICAL Axial Symmetry C1 Axial Rise 5.17 Angular Rotation -100.55
Map-Model Fitting and Refinement Id 1 Refinement Space REAL Refinement Protocol RIGID BODY FIT Refinement Target Overall B Value Fitting Procedure Details
Data Acquisition Detector Type GATAN K2 SUMMIT (4k x 4k) Electron Dose (electrons/Å**2) 40
Imaging Experiment 1 Date of Experiment Temperature (Kelvin) Microscope Model FEI TITAN KRIOS Minimum Defocus (nm) 1500 Maximum Defocus (nm) 2500 Minimum Tilt Angle (degrees) Maximum Tilt Angle (degrees) Nominal CS 2.7 Imaging Mode DARK FIELD Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER Nominal Magnification Calibrated Magnification Source FIELD EMISSION GUN Acceleration Voltage (kV) 300 Imaging Details
EM Software Task Software Package Version PARTICLE SELECTION RELION 2.0 IMAGE ACQUISITION EPU 2.3 CTF CORRECTION RELION 2.0 MODEL FITTING Coot 0.8.9.1 EL MODEL REFINEMENT PHENIX dev 2405 INITIAL EULER ASSIGNMENT RELION 2.0 FINAL EULER ASSIGNMENT RELION 2.0 CLASSIFICATION RELION 2.0 RECONSTRUCTION RELION 2.0
Image Processing CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details NONE 382715