9BH9
Human DNA polymerase theta helicase domain dimer bound to DNA in the microhomology aligning conformation
ELECTRON MICROSCOPY
Sample |
---|
Human DNA polymerase theta helicase domain bound to stem-loop DNA with microhomology in the 3' overhang |
Sample Components |
Human DNA polymerase theta helicase domain |
Stem-loop DNA with microhomology in the 3' overhang |
Specimen Preparation | |
---|---|
Sample Aggregation State | PARTICLE |
Vitrification Instrument | HOMEMADE PLUNGER |
Cryogen Name | ETHANE |
Sample Vitrification Details | 3 microliters of sample was applied to the surface of the grid, blotted for 2 seconds after the liquid spot on the filter paper stopped spreading, and ... |
3D Reconstruction | |
---|---|
Reconstruction Method | SINGLE PARTICLE |
Number of Particles | 107353 |
Reported Resolution (Å) | 3.5 |
Resolution Method | FSC 0.143 CUT-OFF |
Other Details | Non-uniform refinement and local filtering by resolution |
Refinement Type | |
Symmetry Type | POINT |
Point Symmetry | C1 |
Map-Model Fitting and Refinement | |||||
---|---|---|---|---|---|
Id | 1 (5A9J, 2P6R) | ||||
Refinement Space | REAL | ||||
Refinement Protocol | FLEXIBLE FIT | ||||
Refinement Target | Cross-correlation coefficient | ||||
Overall B Value | 67.19 | ||||
Fitting Procedure | |||||
Details | We modeled DNA bases into sharpened density when possible. We modeled the DNA backbone into locally filtered density in low resolution areas. |
Data Acquisition | |||||||||
---|---|---|---|---|---|---|---|---|---|
Detector Type | GATAN K3 BIOQUANTUM (6k x 4k) | ||||||||
Electron Dose (electrons/Å**2) | 50 |
Imaging Experiment | 1 |
---|---|
Date of Experiment | |
Temperature (Kelvin) | |
Microscope Model | FEI TITAN KRIOS |
Minimum Defocus (nm) | 1000 |
Maximum Defocus (nm) | 1000 |
Minimum Tilt Angle (degrees) | |
Maximum Tilt Angle (degrees) | |
Nominal CS | 2.7 |
Imaging Mode | BRIGHT FIELD |
Specimen Holder Model | FEI TITAN KRIOS AUTOGRID HOLDER |
Nominal Magnification | 105000 |
Calibrated Magnification | 60024 |
Source | FIELD EMISSION GUN |
Acceleration Voltage (kV) | 300 |
Imaging Details |
EM Software | ||
---|---|---|
Task | Software Package | Version |
PARTICLE SELECTION | cryoSPARC | 4.3 |
IMAGE ACQUISITION | Leginon | 3.6 |
CTF CORRECTION | cryoSPARC | 4.3 |
MODEL FITTING | UCSF ChimeraX | 1.6.1 |
MODEL FITTING | ISOLDE | 1.6.0 |
INITIAL EULER ASSIGNMENT | cryoSPARC | 4.3 |
FINAL EULER ASSIGNMENT | cryoSPARC | 4.3 |
CLASSIFICATION | cryoSPARC | 4.3 |
RECONSTRUCTION | cryoSPARC | 4.3 |
MODEL REFINEMENT | PHENIX | 1.20.1 |
Image Processing | ||||
---|---|---|---|---|
CTF Correction Type | CTF Correction Details | Number of Particles Selected | Particle Selection Details | |
PHASE FLIPPING AND AMPLITUDE CORRECTION | 5424691 |