☰ Navigation Tabs
EV-D68 in complex with inhibitor Jun11-53-7
Starting Model(s) Initial Refinement Model(s) Type Source Accession Code Details experimental model PDB 7TAG
Sample EV-D68 US/MO/14-18947
Specimen Preparation Sample Aggregation State PARTICLE Vitrification Instrument LEICA EM GP Cryogen Name ETHANE Sample Vitrification Details
3D Reconstruction Reconstruction Method SINGLE PARTICLE Number of Particles 353170 Reported Resolution (Å) 2.3 Resolution Method FSC 0.143 CUT-OFF Other Details Refinement Type Symmetry Type POINT Point Symmetry I
Map-Model Fitting and Refinement Id 1 (7tag) Refinement Space REAL Refinement Protocol OTHER Refinement Target Overall B Value Fitting Procedure Details
Data Acquisition Detector Type GATAN K3 BIOQUANTUM (6k x 4k) Electron Dose (electrons/Å**2) 35.4
Imaging Experiment 1 Date of Experiment Temperature (Kelvin) Microscope Model FEI TITAN KRIOS Minimum Defocus (nm) 600 Maximum Defocus (nm) 2000 Minimum Tilt Angle (degrees) Maximum Tilt Angle (degrees) Nominal CS 2.7 Imaging Mode BRIGHT FIELD Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER Nominal Magnification 64000 Calibrated Magnification Source FIELD EMISSION GUN Acceleration Voltage (kV) 300 Imaging Details
EM Software Task Software Package Version PARTICLE SELECTION cryoSPARC IMAGE ACQUISITION EPU CTF CORRECTION cryoSPARC MODEL FITTING ISOLDE MODEL FITTING UCSF ChimeraX MODEL FITTING Coot MODEL REFINEMENT PHENIX INITIAL EULER ASSIGNMENT cryoSPARC FINAL EULER ASSIGNMENT cryoSPARC CLASSIFICATION cryoSPARC RECONSTRUCTION cryoSPARC
Image Processing CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details PHASE FLIPPING AND AMPLITUDE CORRECTION