human STEAP4 bound to NADP, FAD, heme and Fe(III)-NTA.
Refinement RMS Deviations Key Refinement Restraint Deviation f_dihedral_angle_d 15.4262857146 f_angle_d 1.09614376969 f_chiral_restr 0.0659655032954 f_bond_d 0.00998417181248 f_plane_restr 0.00802478754261
Sample Human Six-Transmembrane Epithelial Antigen of the Prostate 4
Specimen Preparation Sample Aggregation State PARTICLE Vitrification Instrument FEI VITROBOT MARK IV Cryogen Name ETHANE Sample Vitrification Details Blotted for 4 seconds
Blotforce 0
3D Reconstruction Reconstruction Method SINGLE PARTICLE Number of Particles 255144 Reported Resolution (Å) 3.1 Resolution Method FSC 0.143 CUT-OFF Other Details Standard Refinement ant Post processing in Relion. Refinement Type Symmetry Type POINT Point Symmetry C3
Map-Model Fitting and Refinement Id 1 Refinement Space Refinement Protocol Refinement Target Overall B Value Fitting Procedure Details No Fitting was performed.
Data Acquisition Detector Type GATAN K2 SUMMIT (4k x 4k) Electron Dose (electrons/Å**2) 47.4
Imaging Experiment 1 Date of Experiment Temperature (Kelvin) Microscope Model FEI TITAN KRIOS Minimum Defocus (nm) 800 Maximum Defocus (nm) 2000 Minimum Tilt Angle (degrees) Maximum Tilt Angle (degrees) Nominal CS 2.7 Imaging Mode BRIGHT FIELD Specimen Holder Model FEI TITAN KRIOS AUTOGRID HOLDER Nominal Magnification 165000 Calibrated Magnification Source FIELD EMISSION GUN Acceleration Voltage (kV) 300 Imaging Details Grid was prescreened on a 200 kV Talos Arctica Microscope.
Microscope alignment was performed by Wim Hagen at EMBL Heidelberg.
EM Software Task Software Package Version PARTICLE SELECTION RELION 2.1 IMAGE ACQUISITION SerialEM CTF CORRECTION Gctf 1.18 INITIAL EULER ASSIGNMENT RELION 2.1 FINAL EULER ASSIGNMENT RELION 2.1 CLASSIFICATION RELION 2.1 RECONSTRUCTION RELION 2.1
Image Processing CTF Correction Type CTF Correction Details Number of Particles Selected Particle Selection Details PHASE FLIPPING AND AMPLITUDE CORRECTION GCTF within the Relion GUI. 1098075