Find PDB structures and Computed Structure Models (CSM) by combining queries from tools in this suite: Attribute Search, Sequence Similarity, Sequence Motif, 3D Similarity, and 3D Motif with 'AND' logic.
VITRIFICATION 1 -- CRYOGEN- ETHANE, HUMIDITY- 100, TEMPERATURE- 100, INSTRUMENT- OTHER, METHOD- THE SPECIMEN WAS VITRIFIED IN AN ENVIRONMENTAL PLUNGE- ...
VITRIFICATION 1 -- CRYOGEN- ETHANE, HUMIDITY- 100, TEMPERATURE- 100, INSTRUMENT- OTHER, METHOD- THE SPECIMEN WAS VITRIFIED IN AN ENVIRONMENTAL PLUNGE-FREEZE APPARATUS (BELLARE ET AL, J.ELECTR. MICROS. TECH., 1988, 10, 87-111). BLOT FOR 15- 18 SECONDS AFTER THE DIAMETER OF THE BLOTTED MENISCUS CEASES TO EXPAND AND PLUNGED.,
3D Reconstruction
Reconstruction Method
SINGLE PARTICLE
Number of Particles
25492
Reported Resolution (Å)
4.95
Resolution Method
Other Details
SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD -2676(DEPOSITION ID: 12566).AUTHORS HAVE USED COOT TO FIX SOME CLOSE CONTACTS AND RESET THE B-FACT ...
SUBMISSION BASED ON EXPERIMENTAL DATA FROM EMDB EMD -2676(DEPOSITION ID: 12566).AUTHORS HAVE USED COOT TO FIX SOME CLOSE CONTACTS AND RESET THE B-FACTOR. AS THIS MODEL IS NOT REFINED, HENCE THE B-FACTOR DOES NOT MEAN ANYTHING.
Refinement Type
Symmetry Type
POINT
Point Symmetry
C1
Map-Model Fitting and Refinement
Id
1
Refinement Space
REAL
Refinement Protocol
Refinement Target
Overall B Value
Fitting Procedure
Details
Data Acquisition
Detector Type
FEI FALCON II (4k x 4k)
Electron Dose (electrons/Å**2)
64
Imaging Experiment
1
Date of Experiment
2013-10-21
Temperature (Kelvin)
85
Microscope Model
FEI TITAN KRIOS
Minimum Defocus (nm)
2000
Maximum Defocus (nm)
5000
Minimum Tilt Angle (degrees)
Maximum Tilt Angle (degrees)
Nominal CS
2.7
Imaging Mode
BRIGHT FIELD
Specimen Holder Model
Nominal Magnification
47000
Calibrated Magnification
81495
Source
FIELD EMISSION GUN
Acceleration Voltage (kV)
300
Imaging Details
EACH IMAGE WAS EXPOSED FOR 4 SECONDS RESULTING IN A TOTAL DOSE OF ABOUT 64 ELECTRONS.